ENGLISH
•selection of displaying mode;
programming of all the parameters relating to the set-point (if available);
programming of all the parameters relating to the digital filter.
TECHNICAL FEATURES
ACCURACY (VOLTAGE AND CURRENT)
LDI35.AV0: ±0.3% F.S., ±1 dgt (@ 25°C ±5°C, R.H. 60%);
LDI35.AV2:
VDC and ADC: ±0.3% F.S., ±1 dgt (@ 25°C ±5°C, R.H. 60%);
VAC and AAC: ±0.5% F.S., ±1 dgt (@ 25°C ±5°C, R.H. 60%, 50/60Hz,
5 to 100% F.S.).
ACCURACY (LDI35.CF: RTD/ TEMPERATURE)
Pt100/Pt1000: ±0.3% F.S., ±2 dgt (@ 25°C ±5°C, R.H. 60%);
Ni100: ±0.5% F.S., ±2 dgt (@ 25 °C ±5°C, R.H. 60%);
: ±0.3% F.S., ±2 dgt (@ 25°C ±5°C, R.H. 60%).
ACCURACY (LDI35.CF: TC TEMPERATURE)
±0.3% F.S., ±2 dgt (@ 25°C ±5°C, R.H. 60%) from -50°C to the limit of
the input range;
±1% F.S., ±2 dgt (@ 25°C ±5°C, R.H. 60%) from -200°C to -5°C of the
input range.
TEMPERATURE DRIFT (VOLTAGE AND CURRENT)
LDI35.AV0: ±200 ppm/°C;
LDI35.AV2: ±200 ppm/°C.
Gross Automation (877) 268-3700 · www.carlogavazzisales.com · [email protected]